Vertical Flow from a Tank Apparatus Model FM 81

Sci-tech Vertical Flow from A Tank Apparatus Model FM 81 determines these losses at different flow rates.

Item Description

Pressure losses in the flow from tanks are essentially the result of two processes: the jet deflection upon entry into the opening and the wall friction in the opening. As a result of the pressure losses the real discharge is smaller than the theoretical flow rate. Different diameters as well as inlet and outlet contours of the openings can be studied. Additionally, the contraction coefficient can be determined as a characteristic for different contours. The experimental unit includes a transparent tank, a measuring device as well as a pitot tube and twin tube manometers. An interchangeable insert is installed in the tank’s water outlet to facilitate the investigation of various openings. Five inserts with different diameters, inlet contours and outlet contours are provided along with the unit. The issued water jet is measured using a measuring device. A pitot tube detects the total pressure of the flow. The pressure difference (read on the manometer) is used to determine the velocity. The tank is fitted with an adjustable overflow and a measuring point for static pressure. In this way, the level can be precisely adjusted and read on the manometer. The experimental unit is positioned easily and securely on the work surface of the HM 150 base module. The water is supplied and the flow rate measured by HM 150. Alternatively, the experimental unit can be operated by the laboratory supply. The well-structured instructional material sets out  the fundamentals and provides a step-by-step guide through the experiments.

Technical Specifications

Specifications

[1] study of pressure losses in vertical flows from tanks

[2] determining the contraction coefficient for different contours and diameters

[3] tank with adjustable overflow

[4] 5 interchangeable inserts with different contours

[5] measuring device for determining the jet diameter

[6] pitot tube for determining the total pressure

[7] pressure display on twin tube manometers

[8] flow rate determined by HM 150 base module

[9] water supply using HM 150 base module or via laboratory supply.

Technical Specifications:

Tank

– capacity: approx. 13L

– overflow height: max. 400mm

– max. flow rate: 14L/min

Inserts

Inside diameters: d1=inlet, d2=outlet

– 1x cylindrical hole, d=12mm
– 1x outlet from the insert: cone d1=24mm, d2=12mm

– 1x inlet to the insert: orifice plate d1=24mm, d2=12mm

– 1x inlet to the insert: cone d1=30mm, d2=12mm

– 1x inlet to the insert: rounded, d=12mm

Measuring ranges

– pressure: 500mmWC

– jet radius: 0…10mm

Experiments

– study of the outlet jet (diameter, velocity)

– determination of pressure losses and contraction coefficient for different outlet contours

– determination of flow rate at different discharge heads

Services Required
– Mains power supply: 220-240V, 1Ph, 50Hz
– Closed water circuit or ‘Sci-tech’ Hydraulic Bench FM 100.